Based on the customer information submitted, the staff member was cut by the halstead mosquito forcep sm161 that was sticking through the surgical pack on (b)(6) 2017.The staff member went to occupational medicine per hospital procedure and received a tetanus shot as a precautionary measure because he could not recall when he had had one last.All production process systems were reviewed and no discrepancies were found.This wo # (b)(4) was for (b)(4) kits and it was manufactured in lines (b)(6) and (b)(6) on (b)(6) 2017.(b)(4) kits were inspected by the production team with no defects found.(b)(4) kits were inspected by the qa team with no defects found.No rts or a/p for this component.The investigation was completed.The pack design of wo# (b)(4) was evaluated by the design team.The picture provided and the design of the kit were reviewed and we found a risk for the btc # (b)(4) to be punctured by the sharp component items # (b)(4), sm106 & sm161 since they are into a zip lock bag near to the btc.The kit was flagged for design change on (b)(6) 2017 to look for a safer placement for the sharp components and these components will be protected with soft components to avoid any damage to the btc.
|