• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number E2781R-28ASP
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/13/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation: the incident sample was requested but to date has not been received for evaluation.If the sample or additional information pertinent to the incident is obtained, a follow-up report will be submitted.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The customer reported that during an low anterior resection (lar) procedure, electric leakage occurred near the aspiration holes (asp).A new device was opened but also had electrical leakage within one hour of use.A new device was opened again to continue.There was no patient harm.
 
Additional Manufacturer Narrative
Evaluation summary one device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product did not meet specification as received.Visual inspection found the device to have melted around the aspiration holes.The reported condition was confirmed.No insulation was missing from the device.The device failed hipot testing around the damaged insulation indicating electrical leakage.The melting occurred as a result of arcing.Two possible sources of arcing were identified.Simultaneously activating the suction function and the electrical current may result in increased arcing at either the electrode tip or aspiration holes, causing the insulation to melt.Activating the electrode while retracted in the sheath could result in arcing through the aspiration holes and potential melting.The investigation identified the root cause of the reported event to be attributed to the user.The instructions for use (ifu) states, simultaneously activating suction/irrigation and electrosurgical current may result in increased arcing at either the electrode tip or aspiration holes, and/or burns to adjacent tissue.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key7004692
Report Number1717344-2017-06164
Device Sequence Number10998875
Product Code GEI
UDI-Device Identifier20884524000708
UDI-Public20884524000708
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K944244
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/04/2022
Device Model NumberE2781R-28ASP
Device Catalogue NumberE2781R-28ASP
Device Lot Number71130132X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2017
Initial Date Received by Manufacturer 10/13/2017
Supplement Date Received by Manufacturer02/19/2018
Initial Report FDA Received Date11/06/2017
Supplement Report FDA Received Date03/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-