Evaluation: the incident sample was requested but to date has not been received for evaluation.If the sample or additional information pertinent to the incident is obtained, a follow-up report will be submitted.If information is provided in the future, a supplemental report will be issued.
|
|
Correction: evaluation summary: one sonicision cordless ultrasonic dissector was returned for evaluation.The returned product did not meet specification as received.Visual inspection of the disposable hand piece revealed that the waveguide had fractured and the tip had broken off.The broken tip was returned with the device.The reported condition was confirmed.The waveguide was inspected under magnification to identify the point of initial contact that caused the fracture and eventual break.Investigation personnel concluded that the titanium waveguide was in use when it fractured.Further investigation revealed that the dissector tip came in contact with a metal object (hemostats, clips, staples, retractors, etc.) during activation.This contact caused the waveguide to crack and eventually break off.The investigation identified the cause of the reported event to be user error.The device instructions for use currently contain a warning against contact between the dissector tip and metal objects (hemostats, clips, staples, retractors, etc.) during activation.This issue is specific to ultrasonic dissectors.The ifu also advices device users to visually inspect all system components for breaks, cracks, nicks, or other damages prior to use.Ifu also states: do not use damaged components.Use of damaged components may result in injury to the patient or user.If information is provided in the future, a supplemental report will be issued.
|