| Model Number |
48557314 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/10/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; oasys system was used on c3-t1 pcf.Surgeon placed all of his screws and decided to take one out.When he tried backing out the screw, the tulip head broke off.He was able to remove the screw shank a few minutes later.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection; device history review; complaint history review; risk assessment; visual inspection showed the tulip was confirmed separated from the screw shank.Manufacturing records for this product were reviewed and no anomalies were found.The probable root cause is mis alignment of the screwdriver to the screw axis.
|
| |
|
Event or Problem Description
|
|
It was reported that; oasys system was used on c3-t1 pcf.Surgeon placed all of his screws and decided to take one out.When he tried backing out the screw, the tulip head broke off.He was able to remove the screw shank a few minutes later.
|
| |
|