• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. FOUNDATION SHOULDER; GLENOID, UNIFORM, PEGGED 50MM CM FOUNDATION SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. FOUNDATION SHOULDER; GLENOID, UNIFORM, PEGGED 50MM CM FOUNDATION SHOULDER Back to Search Results
Model Number 500-04-150
Health Effect - Clinical Code Pain (1994)
Date of Event 10/09/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient being in pain from loose glenoid component.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the patient was in pain.The in-vivo length of patient service for the implant was 12 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.This event is deemed to be non-product related.The root cause of this complaint was a revision surgery due to the pain because of loose glenoid.There are many factors that may contribute to the event that are outside the control of djo surgical are loose joints from inadequate soft tissue support, degenerative bone, patient bone deterioration, patient activities or trauma.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION SHOULDER
Common Device Name
GLENOID, UNIFORM, PEGGED 50MM CM FOUNDATION SHOULDER
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key7005928
Report Number1644408-2017-00977
Device Sequence Number16426596
Product Code KWS
UDI-Device Identifier00888912143226
UDI-Public(01)00888912143226
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K960906
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2010
Device Model Number500-04-150
Device Catalogue Number500-04-150
Device Lot Number260192
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/09/2017
Supplement Date Received by Manufacturer11/21/2017
Initial Report FDA Received Date11/07/2017
Supplement Report FDA Received Date12/08/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
-
-