• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC ULTRAFLEX M 100BX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC ULTRAFLEX M 100BX Back to Search Results
Model Number 33102
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that a large amount of adhesive was left on the patient's skin following removal of the catheter.The patient reportedly experienced difficulty removing the adhesive and had used a petroleum product and warm compresses to successfully remove the adhesive.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply wash penis with mild soap and warmwater.Dry thoroughly.Trim pubic hair if necessary.Unroll self-adhering catheter over penis.Gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Connect to drainage device.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." (b)(4).
 
Event or Problem Description
It was reported that a large amount of adhesive was left on the patient's skin following removal of the catheter.The patient reportedly experienced difficulty removing the adhesive and had used a petroleum product and warm compresses to successfully remove the adhesive.
 
Additional Manufacturer Narrative
Received 8 unopened ultraflex mecs.The reported event was unconfirmed, since the samples met specification.The samples were visually inspected for holes or damage on the package, legibility of printed numbers and letters, expiration date, correct impression, print color, missing parts of printing, position of printing, damaged component, foreign matter.Per functional evaluation, the samples were within specification, the acceptable peel strength must be in a range from 0.60 to 2.10 lbf.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply wash penis with mild soap and warmwater.Dry thoroughly.Trim pubic hair if necessary.Unroll self-adhering catheter over penis.Gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Connect to drainage device.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." (b)(4).
 
Event or Problem Description
It was reported that a large amount of adhesive was left on the patient's skin following removal of the catheter.The patient reportedly experienced difficulty removing the adhesive and had used a petroleum product and warm compresses to successfully remove the adhesive.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHESTER MEC ULTRAFLEX M 100BX
Common Device Name
MEC
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7007746
Report Number1018233-2017-05752
Device Sequence Number155673
Product Code NNX
UDI-Device Identifier00801741070679
UDI-Public(01)00801741070679
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Patient
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/28/2022
Device Model Number33102
Device Catalogue Number33102
Device Lot NumberJUAX0863
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2017
Initial Date Received by Manufacturer 10/17/2017
Supplement Date Received by Manufacturer11/17/2017
01/18/2018
Initial Report FDA Received Date11/07/2017
Supplement Report FDA Received Date11/21/2017
01/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-