| Model Number |
VT-XXXXX |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information was not provided.The involved device has been requested for return to livanova for evaluation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
| |
|
Event or Problem Description
|
|
Livanova received a report that the vent valve of the malleable vent catheter separated from the device during a procedure.The device was changed out without delay as soon as the issue was detected.There was no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
One malleable vent catheter was returned to livanova for evaluation.Visual inspection identified that the connector was completely detached from the catheter tubing.No evidence of uv adhesive was present at the connection site.This lack of adhesive application was a manufacturing error.The individual responsible for this portion of the catheter assembly was identified based on the reported lot number and the possible causes for this error were discussed.Remedial training was performed to ensure that the proper steps needed for this product were clearly understood.
|
| |
|