Pump passed the displacement test but alarmed compromised force sensor system during the basic occlusion test due to a loose/protruded drive support disk.Unable to perform the occlusion, prime and excessive no delivery test due to the compromised force sensor system alarm.No moisture damage on the lcd board, mother board, interface board, rf board, motor, vibrator motor and battery tube assembly during visual inspection.Pump received with cracked case at the display window corner, broken reservoir tube lip, minor scratched lcd window, stained end cap sticker, stained address/serial number label, broken belt clip slot, cracked case reservoir tube window corner, cracked lcd window, cracked reservoir tube window and cracked battery tube threads.
|