| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the reporter contacted animas, alleging a power (intermittent power) issue.There was no indication that the product caused or contributed to an adverse event.This issue is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 05-dec-2017 with the following findings: a review of the pump¿s black box and alarm history showed evidence of pump reboot events.The battery compartment and returned battery cap were found to be undamaged, free of moisture ingress, and able to properly secure during testing.The pump successfully completed the 24 hour duration test without any power or reset issues.The pump cover was removed and there was no evidence of moisture or damage to the power circuit.The original complaint of a power issue was noted in the pump history but unable to be duplicated during investigation.
|
| |
|