| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the reporter contacted animas, alleging a audio tone/vibration (intermittent sound) issue.It was alleged that the vibratory alarm was working.There was no indication that the product caused or contributed to an adverse event.This issue is being reported because the issue may cause the user to miss an alarm or warning that may cause cessation of insulin delivery.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 15-nov-2017 with the following findings: during investigation, the pump powered up to a clear auditory alarm.The pump was opened to find no intermittent conditions on the piezo.The intermittent sounds complaint was unable to be duplicated.
|
| |
|