• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC NCIRCLE DELTA WIRE TIPLESS STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC NCIRCLE DELTA WIRE TIPLESS STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Catalog Number NTSED-024115-UDH-MB
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/19/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The event is currently under investigation.A follow up report will be submitted upon receipt of additional information or completion of the investigation.
 
Event or Problem Description
The user facility reported that the ncircle delta wire tipless stone extractor basket would not open prior to use on the patient.There was no patient contact with this device.There were no adverse consequences or effects to the patient reported due to this occurrence.Additional patient and event information was requested but not provided.
 
Additional Manufacturer Narrative
Investigation - evaluation: a review of manufacturing instructions, specifications, the actual returned device, quality control data, and drawings was conducted during the investigation.The returned device was found to be fully closed, and the basket could not be opened, confirming the device was non-functional.Observed damage to the sheath was preventing the basket from opening.Current controls are in place in manufacturing to assure device functionality and integrity prior to shipping.Review of production and quality documentation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.There is no indication that a design or process related failure mode contributed to this event.Based on available information, a definitive root cause can not be determined at this time.Per the quality engineering risk assessment, no further action is required.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NCIRCLE DELTA WIRE TIPLESS STONE EXTRACTOR
Common Device Name
FFL DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7008743
Report Number1820334-2017-03914
Device Sequence Number1184040
Product Code FFL
UDI-Device Identifier00827002191112
UDI-Public(01)00827002191112(17)191208(10)7502371
Combination Product (Y/N)N
Initial Reporter CountryKR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberNTSED-024115-UDH-MB
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 10/25/2017
Supplement Date Received by Manufacturer01/25/2018
Initial Report FDA Received Date11/08/2017
Supplement Report FDA Received Date02/20/2018
Patient Sequence Number1
-
-