• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES SPINE EXPEDIUM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES SPINE EXPEDIUM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 186100001
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is available for investigation.Investigation will be conducted.Follow up will be filed with the findings.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The surgeon distracted the growing rod construct and during final tightening, the single innie setscrew stripped out.
 
Additional Manufacturer Narrative
Visual examination of the returned device revealed the threads were peeled and torn.A review of the device history record could not be performed as the lot number was illegible.A trend analysis was conducted.No emerging trends were found requiring further actions.A definitive root cause cannot be positively determined.However, based on the visual evaluation of the returned sample, the possible cause of the defect is the instrument been subjected to excessive force or anticipated torsional forces during use.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
DEPUY SYNTHES SPINE
325 paramount drive
raynham MA 02767
MDR Report Key7008963
Report Number1526439-2017-10942
Device Sequence Number1196932
Product Code NKB
UDI-Device Identifier10705034103846
UDI-Public(01)10705034103846
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K160904
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number186100001
Device Catalogue Number186100001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2017
Initial Date Received by Manufacturer 10/11/2017
Supplement Date Received by Manufacturer11/09/2017
Initial Report FDA Received Date11/08/2017
Supplement Report FDA Received Date11/15/2017
Patient Sequence Number1
-
-