MAUDE Adverse Event Report: DEPUY SYNTHES SPINE EXPEDIUM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES SPINE EXPEDIUM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 179712740
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Failure of Implant (1924); Injury (2348); Device Embedded In Tissue or Plaque (3165)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the findings.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Revision of s1 left.For broken screw.When the surgeon went to remove the screw, the tip broke part of the shank is still in patient could not retrieve, placed a new screw in s2.Surgeon thinks original surgery was done in 2014.
 
Additional Manufacturer Narrative
Visual examination at the macroscopic level revealed that the polyaxial screw fractured along the screw shank.The distal portion of the screw shank was not returned for analysis.Device was sent for fracture analysis which suggests fatigue first propagated along the progression line across the smooth surface region, then full failure / breakage occurred in the rough region at the termination site.No material defects or other abnormalities have been identified in the fracture analysis report.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.A trend analysis was conducted.No emerging trends were found requiring further actions.A definitive root cause for the polyaxial screw fracturing cannot be positively determined.However, the fracture analysis report identified that the failure was indicative of fatigue.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
DEPUY SYNTHES SPINE
325 paramount drive
raynham MA 02767
MDR Report Key7009075
Report Number1526439-2017-10955
Device Sequence Number498266
Product Code NKB
UDI-Device Identifier10705034067896
UDI-Public(01)10705034067896
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K160904
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number179712740
Device Catalogue Number179712740
Device Lot NumberAJFBRV
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2017
Initial Date Received by Manufacturer 10/12/2017
Supplement Date Received by Manufacturer11/13/2017
Initial Report FDA Received Date11/08/2017
Supplement Report FDA Received Date11/15/2017
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-