| Model Number |
45-782040-L |
| Medical Device Problem Codes |
Detachment Of Device Component (1104); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/10/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was provided.A review of the approved device history record indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The coil remains implanted and the remainder of the device was discarded at the user facility; therefore, a product analysis could not be performed.The root cause cannot be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that resistance was encountered during placement of the 6th embolization coil in the distal splenic artery.The coil unexpectedly detached in the microcatheter.The coil was pushed into the intended area with a guidewire.There was no reported patient injury or intervention.The patient is reported to be stable.
|
| |
|