|
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Conclusion: the device was returned for evaluation.A visual inspection found the device returned with a guidewire stuck within the catheter shaft due to stretching noted along the shaft length.Additionally, the device was returned with the proximal segment of the shaft cut from the device, including the hubs; this cut segment was not returned.The device was also found to be inserted through a 9fr introducer sheath.The sheath was able to be move freely on the shaft.The balloon was inspected, and bunching of the balloon material was noted at the distal end of the balloon.Based on the stretching on the catheter shaft, and bunching of the balloon material, the investigation is confirmed for the reported retraction issues.The investigation is inconclusive for the reported folding issue, as the balloon was returned bunched and as the device was not able to be inflated, due to the proximal portion of the device cut from the device.It is possible that a balloon folding improperly in the vessel could contribute to a retraction issues, and the bunching noted on the returned device.However, as the reported folding issue could not be confirmed, based on the returned sample condition, the definitive root cause for the identified retraction issue could not be confirmed.It is unknown if procedural issues contributed to the reported and identified issues.Labeling review: the current ifu (instructions for use) states: warnings: when the catheter is exposed to the vascular system, it should be manipulated while under high-quality fluoroscopic observation.Do not advance or retract the catheter unless the balloon is fully deflated.If resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.Do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.Precautions: if resistance is felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and guidewire/introducer sheath as a single unit.Use of the atlas pta dilatation catheter: apply negative pressure to fully evacuate fluid from the balloon.Confirm that the balloon is fully deflated under fluoroscopy.While maintaining negative pressure and the position of the guidewire, withdraw the deflated dilatation catheter over the wire through the introducer sheath.Use of a gentle counterclockwise motion may be used to help facilitate catheter removal through the introducer sheath.
|