| Model Number |
M001315051 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Perforation (2001)
|
| Date of Event |
10/13/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It is indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed (b)(4).
|
| |
|
Event or Problem Description
|
|
Same case as mfr: 2134265-2017-10735.It was reported that a perforation occurred.A left atrial appendage (laa) closure procedure was being performed.A watchman® access system (was) along with an unspecified pigtail catheter were inserted into the patient's left atrium.The pigtail was removed since it was not able to gain adequate position in the laa.Then an imager ii angiographic catheter was advanced into the was when a minor perforation was observed.The patient remained stable.Heparin was reversed and vitamin k was ordered.The procedure was aborted.No further complications were reported and the patient status is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was further reported that the imager catheter was extended beyond the was.
|
| |
|