Catalog Number 1125350-38
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: the device was returned for analysis.The reported difficult to position, difficult to remove and shaft separation were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that the stent delivery system (sds) met resistance during advancement over the guide wire due to a buildup of procedural contaminants (blood, contrast), such that the inner member bunched; thus, causing the sds to become stuck on the guide wire and resulting in the reported difficult to position and the reported difficult to remove.Attempting to separate the devices using excessive manipulation resulted in the reported shaft/balloon separation and the noted inner member bunching.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that during a procedure of the unspecified vessel, the 3.5x38 mm xience alpine was being advanced over the grand slam guide wire, while still outside of the anatomy, but became stuck on the wire.While attempting to separate the devices using excessive manipulation, the shaft of the xience alpine separated.The devices were not used in the procedure and were removed as a single unit.Reportedly, there was no issue with the grand slam wire.There was no reported adverse patient sequela or a clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7014690
Report Number2024168-2017-08840
Device Sequence Number799431
Product Code NIQ
UDI-Device Identifier08717648199578
UDI-Public08717648199578
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/22/2020
Device Catalogue Number1125350-38
Device Lot Number7061941
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2017
Initial Report FDA Received Date11/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-