• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. ETHILON SUTURE 18"(45CM) 5-0 BLK; SUTURE, NONABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. ETHILON SUTURE 18"(45CM) 5-0 BLK; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number 661H
Device Problem Component Incompatible (1108)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported that the patient underwent an unknown procedure and suture was used.The wrong needle was found on the suture.There were no adverse patient consequences reported.
 
Manufacturer Narrative
(b)(4).Representative sample was returned for analysis.During the visual inspection of sample, no defects were found on the package.The sample was opened and the swage and attachment area were as expected.The suture was dispensed without problems and examined along of the strand and no defects, damage was observed.The returned product code contains a fs-2 needle and the artwork has been reviewed with the needle specification and no deviation were found; all results meet the specified quality requirements and the needle belong to product code.According to sample condition no defects or deviations were observed on the needle and meet the needle specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHILON SUTURE 18"(45CM) 5-0 BLK
Type of Device
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville 08876 0151
Manufacturer (Section G)
ETHICON INC.
road 183, km. 8.3
san lorenzo
Manufacturer Contact
darlene kyle
p.o. box 151, route 22 west
somerville 08876-0151
9082182792
MDR Report Key7015619
MDR Text Key91945097
Report Number2210968-2017-71170
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
PMA/PMN Number
K946173
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Catalogue Number661H
Device Lot NumberLDH881
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received12/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/27/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-