• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTECH SA ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTECH SA ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA ONE
Device Problem Unintended Collision (1429)
Patient Problem No Code Available (3191)
Event Date 10/19/2016
Event Type  Injury  
Manufacturer Narrative
The device rosa one bs16001 is not fda approved, but it is similar to the device rosa brain 3.0, classified haw - k151359.Investigation indicated that the device computer configuration was modified but not set back to the correct configuration before device shipment to customer.
 
Event Description
It was reported that during a surgery, the robot arm collided with the robot stand reference.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA ONE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key7016837
MDR Text Key91793410
Report Number3009185973-2017-00784
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
PN/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Health Professional
Remedial Action Repair
Type of Report Initial
Report Date 11/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2017
Is this an Adverse Event Report? Yes
Device Operator Health Professional
Device Model NumberROSA ONE
Device Catalogue NumberN/A
Device Lot NumberN/A
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/19/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/12/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberN/A
Patient Sequence Number1
-
-