Catalog Number 328506 |
Device Problems
Detachment Of Device Component (1104); Sticking (1597)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/17/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medical device expiration date: n/a.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that a bd ¿ insulin syringe 31g x 8 mm was found during use where the needle had separated from the hub and became stuck in the insulin vile.There was no report of injury or medical intervention needed.
|
|
Manufacturer Narrative
|
Bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.A review of the manufacturing records was completed for the incident lot and no issues were identified.Conclusion: without a sample, an absolute root cause for this incident cannot be determined as bd was not able to duplicate or confirm the customer¿s indicated failure mode.
|
|