| Catalog Number |
02.09.0117H |
| Medical Device Problem Code |
Device Issue (2379)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This is the second revision surgery underwent by the patient.The patient had a primary knee on (b)(6) 2014.During the primary the torque driver was used.The patient came in due to signs of infection.The patient tested positive for granulicatella adiacens.On (b)(6) 2016 the surgeon removed all implants.On (b)(6) 2016 the surgeon re-implanted medacta components.In (b)(6) 2017 the surgeon determined that the screw and hinge mechanism dissociated from the morse taper of the yoke.The surgeon planned to perform a poly swap on (b)(6) 2017.Batch review performed on 09 november 2017.Lot 140422: (b)(4) items manufactured and released on 19 february 2014.Expiration date: 2019-01-31.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
|
| |
|
Event or Problem Description
|
|
The patient originally came in due to a clicking noise in her knee.The surgeon determined the screw had backed out of the yoke.The surgeon performed a poly swap.There was no damage to the screw or the yoke.The surgeon threaded the screw back into the yoke.It went in with no issue.The surgery was completed successfully.
|
| |
|