• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD CAPROSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD CAPROSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID Back to Search Results
Model Number XC-845
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/13/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra operatively, while closing the skin on suture broke.Another suture was used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted one partial suture and one needle.The needle was observed to be attached to the suture.As no sealed products were returned, functional testing was precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPROSYN
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7018075
Report Number9612501-2017-06556
Device Sequence Number16563198
Product Code GAM
UDI-Device Identifier10884521113558
UDI-Public10884521113558
Combination Product (Y/N)N
Initial Reporter CountrySF
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 01/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Model NumberXC-845
Device Catalogue NumberXC-845
Device Lot NumberD6E1732X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2017
Initial Date Received by Manufacturer 10/13/2017
Supplement Date Received by Manufacturer01/23/2018
Initial Report FDA Received Date11/10/2017
Supplement Report FDA Received Date01/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-