• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA MECTALIF OBLIQUE INTERBODY FUSION DEVICE PEEK/TI 12X28X7 L0°

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA MECTALIF OBLIQUE INTERBODY FUSION DEVICE PEEK/TI 12X28X7 L0° Back to Search Results
Catalog Number 03.26.034
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Mectalif posterior handle (ref.03.22.10.0056.Lot unknown), mectalif oblique handle (ref.03.22.10.0262, lot unknown) and mectalif inner rod (ref.03.22.10.0055, lot unknown) were used during surgery.On (b)(6) 2017 the r and d project manager performed a visual inspection of the retrieved items and commented as follows: both the posterior and the oblique cages present damaging of the coating and scratches particularly in the back side of the devices.The posterior cage is not deformed and the threaded interface is functional.The oblique cage is deformed but the threaded interface is functional.The scratches on the back of the devices are compatible with a rotation of the implants after the fixation on the inserter.The application of a rotation movement with an high force or a not perfect fixation of the devices on the inserter can produce this kind of scratches.The damaging of the coating of both cages and the deformation of the oblique one are probably due to the removal process.Batch review performed on 09 november 2017.Mectalif posterior interbody fusion device peek/ti 11x25x7 l5°, code 03.27.020, lot.148876 (k133192) (b)(4).
 
Event or Problem Description
It was hard to insert the cages into the intervertebral at l4/l5, even though the surgeon tried to trial implant and distracted the dedicated size.After that the surgeon found the cage deformation.There was no delay as they used backup implants to proceed.The surgeon's comments: there is a problem of the implant inserter itself.When hitting an inserter with a hammer, the inner part was loosen.As a result, when inserting the cage into the intervertebral, appropriate force would not be added to the appropriate place.It would be the root cause of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MECTALIF OBLIQUE INTERBODY FUSION DEVICE PEEK/TI 12X28X7 L0°
Common Device Name
OBLIQUE INTERBODY FUSION DEVICE
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
MDR Report Key7018102
Report Number3005180920-2017-00651
Device Sequence Number1005646
Product Code MAX
UDI-Device Identifier07630030830549
UDI-Public07630030830549
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133192
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/03/2021
Device Catalogue Number03.26.034
Device Lot Number1620018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/12/2017
Initial Report FDA Received Date11/10/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age45 YR
Patient Weight103
-
-