• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. TURON SHOULDER; TURON PEGGED GLENOID IMPLANT, SIZE 50

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. TURON SHOULDER; TURON PEGGED GLENOID IMPLANT, SIZE 50 Back to Search Results
Model Number 520-01-250
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/23/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient having a loose glenoid component.
 
Additional Manufacturer Narrative
The reason for this revision surgery is due to a loose glenoid.The previous surgery and the revision detailed in this investigation occurred 6 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to a loose glenoid.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's loosening.There are many factors that may contribute to the event that are outside the control of djo surgical such as: loose joints from inadequate soft tissue support, degenerative bone, patient bone deterioration, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
TURON PEGGED GLENOID IMPLANT, SIZE 50
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7018660
Report Number1644408-2017-00997
Device Sequence Number1423816
Product Code KWS
UDI-Device Identifier00888912024976
UDI-Public(01)00888912024976
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K123982
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2016
Device Model Number520-01-250
Device Catalogue Number520-01-250
Device Lot Number193G1013
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/23/2017
Supplement Date Received by Manufacturer11/22/2017
Initial Report FDA Received Date11/10/2017
Supplement Report FDA Received Date12/11/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
-
-