| Catalog Number |
RBY4C0835 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/16/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The hospital discarded the device.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure using ruby coils.During the procedure, while attempting to advance a ruby coil through a lantern delivery microcatheter (lantern), the physician determined that a ruby coil was the incorrect size and decided to remove and re-sheath it.However, while re-sheathing the ruby coil outside the patient¿s body, the ruby coil detached.The procedure was completed using other ruby coils and the same lantern.There was no report of an adverse effect to the patient.
|
| |
|