The biomedical engineer reported that the org was frequently going into signal loss with two telemetry devices.After some troubleshooting, it was decided that they would send the device in for repair.It is currently awaiting evaluation.The device was in use on patients, however no patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
The biomedical engineer reported that the org was frequently going into signal loss with wo telemetry devices.After some troubleshooting, it was decided that they would send the device in for repair.The device was in use on patients, however, no patient harem was reported.During evaluation of the device, the reported malfunction could not be duplicated.The unit completed four days of extensive testing.Unit is operating within manufacturer's specifications.
|