Model Number 0168L20
Health Effect - Clinical Code Foreign Body In Patient (2687)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the catheter balloon burst within the patient and pieces were missing.A cystoscopy was performed to retrieve the pieces.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: "to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient" (b)(4).
 
Event or Problem Description
It was reported that the catheter balloon burst within the patient and pieces were missing.A cystoscopy was performed to retrieve the pieces.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER COUDE TIP
Common Device Name
COUDE CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7024269
Report Number1018233-2017-05821
Device Sequence Number2037045
Product Code EZC
UDI-Device Identifier00801741017841
UDI-Public(01)00801741017841
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0168L20
Device Catalogue Number0168L20
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/25/2017
Supplement Date Received by Manufacturer11/21/2017
Initial Report FDA Received Date11/13/2017
Supplement Report FDA Received Date11/23/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-