• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOSPIRA COSTA RICA LTD. PLUM 360 DRIVER NEW; PUMP INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOSPIRA COSTA RICA LTD. PLUM 360 DRIVER NEW; PUMP INFUSION Back to Search Results
Catalog Number 30010
Medical Device Problem Codes Thermal Decomposition of Device (1071); Device Emits Odor (1425)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/23/2017
Type of Reportable Event Malfunction
Event or Problem Description
The event involved a customer allegation that there was a burnt board and it smelled like burned plastic or an electronic component.During visual inspection, fse testing confirmed burnt components were found on the power supply and determined the probable cause was damage to the power supply board.There were no reports of any adverse patient events or delays in critical therapies, while the device was in clinical use.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUM 360 DRIVER NEW
Common Device Name
PUMP INFUSION
Manufacturer (Section D)
HOSPIRA COSTA RICA LTD.
9615050
zona franca global,
la aurora heredia
CS 
MDR Report Key7025789
Report Number9615050-2017-00130
Device Sequence Number11095771
Product Code FRN
UDI-Device IdentifierM3353001004051
UDI-Public+M3353001004051$$+70021774013/16D20161006
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K141789
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number30010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2017
Initial Report FDA Received Date11/13/2017
Patient Sequence Number1
-
-