ENCORE MEDICAL L.P. BIPOLAR HIP; SHELL, BIPOLAR, MODULAR, ENCORE, SIZE56 Back to Search Results
Model Number 412-02-056
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/24/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient having bipolar head protrusion, so the surgeon removed the bipolar and replaced with a cage.The cage was not sufficient, so he replaced with a 60 bipolar head.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the patient had bipolar head protrusion.The in-vivo length of patient service for the implant was 22 days apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was kept at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.This event is deemed to be non-product related.The root cause of this complaint was a revision surgery due to the bipolar head protrusion.There are multiple factors that may contribute to the event that are outside the control of djo surgical are patient activities, excessive loads or torques, excessive range of motion, patient bone deterioration or trauma.Due to short time between the previous and revision surgery, it is also possible that the event may have occurred due to incorrect implant selection, improper surgical procedure or patient noncompliance with medical instructions.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOLAR HIP
Common Device Name
SHELL, BIPOLAR, MODULAR, ENCORE, SIZE56
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key7026259
Report Number1644408-2017-01000
Device Sequence Number11232796
Product Code KWY
UDI-Device Identifier00888912077460
UDI-Public(01)00888912077460
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K953510
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/08/2022
Device Model Number412-02-056
Device Catalogue Number412-02-056
Device Lot Number004C1068
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/24/2017
Supplement Date Received by Manufacturer11/28/2017
Initial Report FDA Received Date11/13/2017
Supplement Report FDA Received Date12/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
-
-