| Catalog Number |
9735018 |
| Medical Device Problem Codes |
Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/17/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information not provided due to (b)(6) patient privacy regulations.The suspect suction has been received by manufacturer for evaluation.When the received instrument was connected to a good known system, the suction was recognized but would not track returning a red status.A check of the em interface revealed a issue with coil causing the poor signal.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that during transnasal endoscopic pituitary surgery, the malleable suction was successfully registered at the beginning of the procedure however during the case the suction had recognition failure.A different malleable suction was used to complete the procedure.There was no delay to procedure.No impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
| |
|