| Model Number |
70900-01 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been requested back for investigation.A follow- up report will be submitted once additional information is obtained.Note: the meter is designed to report readings of 20 mg/dl to 500 mg/dl.It should be noted that this meter does not give numeric value for readings higher than 500 mg/dl or lower than 20 mg/dl.A "hi" display message indicates a reading higher than 500 mg/dl and a "lo" display message indicates a reading lower than 20 mg/dl.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter within 10 minutes.Results of "hi" ( >500 mg/dl) and "lo" ( <20 mg/dl) were plotted on the parkes error grid.The results fell into the "d" zone showing the difference in values to be clinically significant.There was no report of death, serious injury or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported product's were returned and investigated.The complaint was not confirmed and no new issues were observed.All results were within range specification and no errors were observed during control solution testing.(meter serial number) was incorrectly documented in the initial mdr 30 day report.Meter serial number has been updated.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter within 10 minutes.Results of "hi" ( >500 mg/dl) and "lo" ( <20 mg/dl) were plotted on the parkes error grid.The results fell into the "d" zone showing the difference in values to be clinically significant.There was no report of death, serious injury or mistreatment associated with this event.
|
| |
|
Search Alerts/Recalls
|