In lieu of a reported lot number, a ship history report (shr) was generated for item number h965750335 in order to determine the last 3 lots shipped to the reporting hospital in the 6 months prior to the procedure date.A review of the device history records was performed for the packaging and component lots generated by the shr for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The recent angiodynamics complaint report was reviewed for the bioflo picc product family for the failure mode "hub broken/cracked." no adverse trends were indicated.The complaint was confirmed, as the female luer lock component of the white valve was seen to be cracked just adjacent to the molded parting line.There were no other visual defects noted to the hub or catheter.A possible root cause for the crack is an over-tightened connection with a mating male luer (likely a needleless connector).Inadequate flushing of the device may also be a contributing factor.The directions for use (dfu) supplied with the reported picc device contain caution that "repeated overtightening may reduce hub connector life," and not to use hemostats to "secure or remove devices with luer lock hub connections." recommended flushing techniques are also stated in the dfu.(b)(4).
|