• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS ANGIODYANMICS / BIOFLO; PERIPHERALLY INSERTED CENTRAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS ANGIODYANMICS / BIOFLO; PERIPHERALLY INSERTED CENTRAL CATHETER Back to Search Results
Catalog Number H965750335
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/19/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
In lieu of a reported lot number, a ship history report (shr) was generated for item number h965750335 in order to determine the last 3 lots shipped to the reporting hospital in the 6 months prior to the procedure date.A review of the device history records was performed for the packaging and component lots generated by the shr for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The recent angiodynamics complaint report was reviewed for the bioflo picc product family for the failure mode "hub broken/cracked." no adverse trends were indicated.The complaint was confirmed, as the female luer lock component of the white valve was seen to be cracked just adjacent to the molded parting line.There were no other visual defects noted to the hub or catheter.A possible root cause for the crack is an over-tightened connection with a mating male luer (likely a needleless connector).Inadequate flushing of the device may also be a contributing factor.The directions for use (dfu) supplied with the reported picc device contain caution that "repeated overtightening may reduce hub connector life," and not to use hemostats to "secure or remove devices with luer lock hub connections." recommended flushing techniques are also stated in the dfu.(b)(4).
 
Event or Problem Description
As reported, picc which had been in patient's arm was removed and replaced on the third day of use because of a cracked hub.No patient injury was reported.The used catheter was returned to angiodynamics for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIODYANMICS / BIOFLO
Common Device Name
PERIPHERALLY INSERTED CENTRAL CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key7032775
Report Number1317056-2017-00098
Device Sequence Number19369292
Product Code LJS
UDI-Device IdentifierH965750335
UDI-PublicH965750335
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K163452
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 11/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other Health Care Professional
Device Catalogue NumberH965750335
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2017
Initial Report FDA Received Date11/15/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age29 YR
-
-