The distal shaft of the catheter was found kinked.Thus, the inner lumen of the catheter did not allow guidewire advancement during lab analysis.A review of same lot history record identified no manufacturing nonconformities.Additionally, a review of the complaint history identified, no other incident from same lot.Based on the information reviewed, there is no indication of product quality issue with respect to manufacturing, design and labelling.Please note: this is the first electronic report being filed, and it was delayed due to the sign up process with the esg webtrader system.
|
|
Mozectm nc rx percutaneous transluminal coronary angioplasty (ptca) 2.50mm x 28mm was tracked over a non-cordis wire and inflated at twenty atmospheres (20atm).The balloon initially tacked well over the wire but after multiple inflations, when the user was pulling the balloon out, the wire got stuck into the balloon.The balloon and the wire were removed together and the balloon was rewired.The procedure was conducted for a chronic total occlusion (cto) of the left anterior descending (lad).No patient injury has been reported.
|