MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problems No Display/Image (1183); Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 10/19/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative inspected the navigation system onsite and confirmed the reported event.The monitor failed visual inspection.The surgeon monitor was replaced and the issue resolved.A full system check-out was completed following part replacement and all tests passed.Full system functionality was confirmed and the system was returned to service.The suspect monitor was returned to the manufacturer for evaluation.After performing functional testing and visual/physical examination, the issue was confirmed as an lcd monitor malfunction.The monitor displayed a good image for about 10 second then went blank.Hot plugging the monitor brought back the image for about 10 seconds before going blank again.
 
Event Description
A medtronic representative reported that outside of a procedure, while testing the navigation system, the surgeon monitor would turn off 10 seconds after plugging in.The representative reseated the surgeon monitor cable and the monitor would turn on, but then turn off again 10 seconds later.There was no patient present when this issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stephanie riley
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7034382
MDR Text Key92858326
Report Number1723170-2017-04685
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 11/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received10/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-