| Catalog Number |
03.010.048 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/23/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device used for treatment, not diagnosis.Patient id, dob & weight is not available for reporting.Device is an instrument and is not implanted/explanted.Device is not expected to be returned for manufacturer review/investigation.Reporters phone number (b)(6).Device history records review was conducted.The report indicates that the: dhr review: part number: synthes lot number: 5759405.Supplier lot number: na.Release to warehouse date: 10-apr-2008.Manufactured by: (b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this products, and any subcomponents, which would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that multiple instruments malfunctioned during a lateral entry femoral nailing to repair a midshaft femur fracture due to a motor vehicle accident on (b)(6) 2017.After the surgeon inserted the nail, he inserted the three-part trocar assembly, he then checked to make sure it was being targeted correctly.When he went to insert the guide wire, it kept hitting the nail and was misaligned.The misalignment issue only happened with one guide wire and first set of the three-part trocar assembly.Later in the procedure, he attempted to take the nail out, but the hammer kept getting jammed and wouldn¿t rock back and forth on the hammer guide, he struggled to get the nail out.Then, as the depth gauge was being used to measure the length of the recon screws; the surgeon tried three times and the measurement was inaccurate, it was 10mm longer than needed.There was a reported surgical delay of 10 minutes to take the nail out and reposition it.No additional x-rays or other medical intervention required.The procedure was completed successfully with the patient in stable condition.This complaint involves nine (5) devices.Concomitant devices reported: hammer guide (part# unknown, lot# unknown, quantity 1), nail (part# unknown, lot# unknown, quantity 1), extraction screw (part# unknown, lot# unknown, quantity 1).This report is 1 of 5 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The recon locking aiming arm for lateral entry femoral nails-ex (part #: 03.010.048, lot #: 5759405) was reported as misaligned.The returned device was investigated and this complaint is unconfirmed.A visual inspection under 5x magnification, functional test, and drawing review were performed.No product design issues or discrepancies were observed.Device was received showing minor wear but nothing that would inhibit functionality.A function test revealed that the aiming arm mated as expected with the returned insertion handle and triple trocar assemblies (03.010.075, 03.010.076 and 03.010.077).The nail was not returned so alignment to the nail could not be replicated.Aiming arm assembly drawing was reviewed during this investigation.No product design issues or discrepancies were observed.A dimensional inspection was not performed because all returned devices assembled as intended.Changed from no to yes.Device returned december 01, 2017.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|