| Catalog Number |
K6412 SMN10445104 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/17/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer contacted siemens healthcare diagnostics concerning the impact of biotin on the tsh result.Siemens ongoing investigation has identified the following new information regarding potential biotin interference, biotin interference limits are currently incorrectly listed in the instructions for use (ifu) for dimension vista thyroid stimulating hormone (tsh) flex reagent cartridges k6412 non-interference information section.The current biotin non-interference ifu information for tsh indicates that the amount of biotin that causes <= 10% interference is 500 ng/ml (2050 nmol/l).Customers were informed that concentrations of biotin above 100 ng/ml (409 nmol/l) can potentially result in interference > 10%.Siemens issued an urgent medical device correction (umdc) communication vc17-04.B.Us dated october 16, 2017 to all u.S.Dimension vista® accounts and an urgent field safety notice (ufsn) vc-17-04.B.Ous, to all outside u.S.Dimension vista® accounts.
|
| |
|
Event or Problem Description
|
|
A discordant depressed thyroid stimulating hormone (tsh) result was obtained on a patient sample on the dimension vista 1500 system.The initial result was reported to the physician and questioned.The same patient sample was tested at an alternate facility with an alternate methodology and a higher result was obtained.The tsh result obtained on the alternate methodology was reported to the physician because the result correlated with the clinical context of the patient.There are no reports of patient intervention or adverse health consequences as a result of the discordant depressed tsh result.
|
| |
|
Additional Manufacturer Narrative
|
|
On a letter dated 28 july 2017 siemens healthcare diagnostics communicated to agence nationale de securite du medicament (ansm- reference number (b)(4)) the following for tsh: the tsh assay has been tested for potential interference at biotin levels of 100, 250, 500, and 1200 ng/ml.Interference testing was performed following clinical laboratory standards institute (clsi) guidance at concentrations approximately equivalent to the levels listed in the ifu.Below is the biotin interference of dimension vista tsh: test level: 4.24, units: iu/ml.Biotin test level and % interference observed: 1200 ng/ml -99.1%, 500 ng/ml -38.7%, 250 ng/ml -11.7%, 100 ng/ml -2.4%.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial mdr 2517506-2017-00826 was filed 11/15/2017.Mdr supplement 2517506-2017-00826 supplement 1 was filed 11/16/2017.Based on input from the (b)(4) recall coordinator on 12/18/2017 this recall was resubmitted to the (b)(4) office (oradevices1recalls@fda.Hhs.Gov) under 2432235-12/19/2017-004-c.Recall changed from 2517506-03/30/2017-003-c to 2432235-12/19/2017-004-c.
|
| |
|