• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DORADO PTA BALLOON DILATATION CATHETER; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DORADO PTA BALLOON DILATATION CATHETER; CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number DR80610
Medical Device Problem Codes Break (1069); Leak/Splash (1354)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/24/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that when flushing the device during preparation for an angioplasty procedure, an alleged leak was observed from the pta balloon catheter shaft.There was no patient contact.
 
Additional Manufacturer Narrative
The device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.The device was returned.A visual inspection found a partial break at the butt joint between the catheter and proximal end of the balloon.An inflation attempt was made and water was seen exiting the break.Therefore, the investigation is confirmed for the reported leak as well as for the identified break at the butt joint.The break at the glue butt joint resulted in the identified leak.However, the definitive root cause for the partial break could not be determined based upon available information.It is unknown whether procedural issues contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings: if resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.Use of the dorado pta dilatation catheter: advance the catheter through the introducer sheath and over the wire to the site of inflation.If the stenosis cannot be crossed with the desired dilatation catheter, use a smaller diameter catheter to pre-dilate the lesion to facilitate passage of a more appropriately sized dilatation catheter.Position the balloon relative to the lesion to be dilated, ensure the guidewire is in place and inflate the balloon to the appropriate pressure.(b)(4).
 
Event or Problem Description
It was reported that when flushing the device during preparation for an angioplasty procedure, an alleged leak was observed from the pta balloon catheter shaft.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO PTA BALLOON DILATATION CATHETER
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7035979
Report Number2020394-2017-01570
Device Sequence Number1745098
Product Code DQY
UDI-Device Identifier00801741058837
UDI-Public(01)00801741058837
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/15/2020
Device Catalogue NumberDR80610
Device Lot Number93SB0088
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2017
Initial Date Received by Manufacturer 10/20/2017
Supplement Date Received by Manufacturer11/27/2017
Initial Report FDA Received Date11/15/2017
Supplement Report FDA Received Date12/08/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-