| Model Number |
010112 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that the coude tip was bent more than usual; therefore, the patient could not insert the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Received 2 unopened urological catheters.The reported event was unconfirmed since the problem could not be reproduced.Per visual inspection, the samples were examined and found the tips were not too curved.The specification for curvature is that the tips not be straight, and they were not.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "visually inspect the product for any imperfections or surface deterioration prior to use." (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the coude tip was bent more than usual; therefore, the patient could not insert the catheter.
|
| |
|
Search Alerts/Recalls
|