Catalog Number FS-OMNI-ACRO-35-260
Device Problem Peeled/Delaminated (1454)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/30/2017
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved could not be reviewed because a lot number was not provided.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use instruct the user to do the following: "for best results, wire guide should be kept wet." failure to flush the wire guide can result in damage to the wire guide.The instructions for use precaution the user that this product is not compatible with metal tip devices."if preloaded, use of wire guide with metal tip ercp devices may result in damage to external coating and/or tip of wire guide." the reported observation can occur if the wire guide was used with an incompatible accessory device.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.Prior to distribution, all fusion preloaded with acrobat wire guide sphincterotomes are subjected to a visual inspection to ensure device integrity.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook fusion pre-loaded with acrobat wire guide.The distal portion of the wire guide coating stripped off.The coating did not fully separate from the wire.Additional information was received on 11/1/2017: it was at the distal/patient end of the wire, roughly 10cm from the tip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE
Type of Device
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key7041113
MDR Text Key93338136
Report Number1037905-2017-00701
Device Sequence Number1
Product Code KNS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFS-OMNI-ACRO-35-260
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/30/2017
Event Location Hospital
Initial Date Manufacturer Received 10/30/2017
Initial Date FDA Received11/17/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No