| Model Number |
CQF7564 |
| Medical Device Problem Codes |
Deflation Problem (1149); Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/21/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device was returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that the pta balloon catheter allegedly was difficult to deflate.There were no reported retraction difficulties through the introducer sheath.Reportedly, the procedure was concluded upon removal of the balloon catheter.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A manufacturing review was performed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.A visual inspection found no deflation related defects with the device.The returned sample was inflated and deflated without any issues within the specified time constraints.Therefore, the investigation is unconfirmed for deflation issues as the balloon functioned as intended during functional testing.Additionally, the device was able to be retracted through an in-house 6fr cordis introducer sheath without issue.Therefore, the investigation is unconfirmed for the reported retraction issues.The definitive root cause for the reported deflation issue could not be determined based upon available information.It is unknown whether patient and/or procedural issues contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings: do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.When the catheter is exposed to the vascular system, it should be manipulated while under high-quality fluoroscopic observation.Do not advance or retract the catheter unless the balloon is fully deflated.If resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.Precautions: if resistance is felt during post procedure withdrawal of the catheter through the introducer sheath, determine if contrast is trapped in the balloon with fluoroscopy.If contrast is present, push the balloon out of the sheath and then completely evacuate the contrast before proceeding to withdraw the balloon.If resistance is still felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and guidewire/introducer sheath as a single unit.
|
| |
|
Event or Problem Description
|
|
It was reported that the pta balloon catheter allegedly was difficult to deflate and subsequent retraction issues through the sheath were encountered.Reportedly, the procedure was concluded upon removal of the balloon catheter.There was no reported patient injury.
|
| |
|