Model Number CQF7564
Medical Device Problem Codes Deflation Problem (1149); Retraction Problem (1536)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/21/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device was returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the pta balloon catheter allegedly was difficult to deflate.There were no reported retraction difficulties through the introducer sheath.Reportedly, the procedure was concluded upon removal of the balloon catheter.There was no reported patient injury.
 
Additional Manufacturer Narrative
A manufacturing review was performed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.A visual inspection found no deflation related defects with the device.The returned sample was inflated and deflated without any issues within the specified time constraints.Therefore, the investigation is unconfirmed for deflation issues as the balloon functioned as intended during functional testing.Additionally, the device was able to be retracted through an in-house 6fr cordis introducer sheath without issue.Therefore, the investigation is unconfirmed for the reported retraction issues.The definitive root cause for the reported deflation issue could not be determined based upon available information.It is unknown whether patient and/or procedural issues contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings: do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.When the catheter is exposed to the vascular system, it should be manipulated while under high-quality fluoroscopic observation.Do not advance or retract the catheter unless the balloon is fully deflated.If resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.Precautions: if resistance is felt during post procedure withdrawal of the catheter through the introducer sheath, determine if contrast is trapped in the balloon with fluoroscopy.If contrast is present, push the balloon out of the sheath and then completely evacuate the contrast before proceeding to withdraw the balloon.If resistance is still felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and guidewire/introducer sheath as a single unit.
 
Event or Problem Description
It was reported that the pta balloon catheter allegedly was difficult to deflate and subsequent retraction issues through the sheath were encountered.Reportedly, the procedure was concluded upon removal of the balloon catheter.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST 40 PTA DILATATION CATHETER
Common Device Name
PTA BALLOON CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7041290
Report Number2020394-2017-01587
Device Sequence Number1745159
Product Code LIT
UDI-Device Identifier00801741060441
UDI-Public(01)00801741060441
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K120660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Model NumberCQF7564
Device Catalogue NumberCQF7564
Device Lot NumberREBU0522
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2017
Initial Date Received by Manufacturer 10/23/2017
Supplement Date Received by Manufacturer01/17/2018
Initial Report FDA Received Date11/17/2017
Supplement Report FDA Received Date01/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR