• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED IMMULITE/IMMULITE 1000 BR-MA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED IMMULITE/IMMULITE 1000 BR-MA Back to Search Results
Model Number IMMULITE/IMMULITE 1000 BR-MA
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/26/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens is evaluating the issue.
 
Event or Problem Description
Siemens technical operations internal investigation has identified that discordant falsely low results were obtained on two investigational samples on an immulite/ immulite 1000 instrument when using the br-ma (ca15-3) assay.The two investigational samples were spiked with biotin at multiple concentrations and run for investigational purposes on an immulite/immulite 1000 instrument.The results of the two samples spiked with 500 ng/ml of biotin were falsely low with a bias of -71% and -51% respectively compared to results of the two samples without biotin.There were no patient samples impacted as this was for investigational purposes only.
 
Additional Manufacturer Narrative
The initial mdr 2432235-2017-00617 was filed on 17-nov-2017.Additional information (06-dec-2017): siemens healthcare diagnostics has confirmed that biotin interferes with the ca15-3 (br-ma) on the immulite/immulite 1000 platform.At biotin concentrations of >100 ng/ml a negative bias that exceeds 10% has been observed on immulite/immulite 1000 platforms.This issue affects previously manufactured, in-date, and future lots.The immulite/immulite 1000 platform instructions for use (ifu's) currently do not have biotin listed as a potential interferent.An urgent field safety notice (ufsn) imc 18-02.A.Ous was sent out to ous customers and an urgent medical device recall (umdr) imc18-02.A.Us was sent to us customers on december 13, 2017.The ufsn and umdr informs the customers of biotin interference on all affected immulite assays.Siemens will update all ifu's with the appropriate biotin information.
 
Additional Manufacturer Narrative
The initial mdr was filed on november 17, 2017.Follow up mdr was filed on december 14, 2017.Corrected information (12/18/2017): based on input from the philadelphia recall coordinator on 12/18/2017 this recall was re-submitted to the (b)(6) office (b)(4).The crr reporting number indicated is incorrect and was changed to (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMMULITE/IMMULITE 1000 BR-MA
Common Device Name
IMMULITE/IMMULITE 1000 BR-MA
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED
glyn rhonwy
llanberis, caernarfon, gwynedd LL55 4EL
UK  LL55 4EL
MDR Report Key7041476
Report Number2432235-2017-00617
Device Sequence Number118538
Product Code MOI
UDI-Device Identifier00630414963655
UDI-Public00630414963655
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K013984
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,study
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIMMULITE/IMMULITE 1000 BR-MA
Device Catalogue NumberLKBR1
Device Lot Number313
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/26/2017
Supplement Date Received by Manufacturer12/06/2017
12/18/2017
Initial Report FDA Received Date11/17/2017
Supplement Report FDA Received Date12/14/2017
12/19/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2432235-12/19/2017-004-C
Patient Sequence Number1
-
-