| Model Number |
1403US |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/24/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.
|
| |
|
Event or Problem Description
|
|
It was reported that the controller had a loose power port connection.The controller was exchanged.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: it was reported that the controller had a loose power port connection.The controller ((b)(4)) was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Review of the manufacturing documentation confirmed that the associated device met all requirements for release.Failure analysis of the returned device revealed that the device passed functional testing.The reported event was confirmed, visual inspection revealed a loose power port 1 connector.Based on an investigation conducted, the most likely root cause of the reported event can be attributed to inadequate thread lock, an inconsistent thread lock cure time and an inadequate torque application during the assembly process.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|