HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1403US
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/24/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.
 
Event or Problem Description
It was reported that the controller had a loose power port connection.The controller was exchanged.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Conclusion: it was reported that the controller had a loose power port connection.The controller ((b)(4)) was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Review of the manufacturing documentation confirmed that the associated device met all requirements for release.Failure analysis of the returned device revealed that the device passed functional testing.The reported event was confirmed, visual inspection revealed a loose power port 1 connector.Based on an investigation conducted, the most likely root cause of the reported event can be attributed to inadequate thread lock, an inconsistent thread lock cure time and an inadequate torque application during the assembly process.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key7041727
Report Number3007042319-2017-04092
Device Sequence Number2037138
Product Code DSQ
UDI-Device Identifier00888707000161
UDI-Public00888707000161
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/31/2015
Device Model Number1403US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2017
Initial Date Received by Manufacturer 10/24/2017
Supplement Date Received by Manufacturer02/20/2018
Initial Report FDA Received Date11/17/2017
Supplement Report FDA Received Date02/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-