| Model Number |
EVOLUTPRO-23-US |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/26/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device remains implanted, therefore no product analysis can be performed.Without return of the device, a conclusive cause cannot be determined.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, the valve was fully deployed and the delivery catheter system (dcs) was removed.The non-medtronic guidewire was then removed from the ventricle and caught on the inflow portion of the valve frame, causing the valve to then dislodged into the ascending aorta.Subsequently, a second transcatheter bioprosthetic valve was implanted.No adverse patient effects were reported.
|
| |
|