Model Number PED2-425-20
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/04/2017
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for analysis as it was implanted in the patient.Since the device was not returned, we are unable to perform further root cause analysis.Per the instructions for use, choose a embolization device with labeled diameter that approximates the target vessel diameter.Select an appropriately sized embolization device such that its fully expanded diameter is equivalent to that of the largest target vessel.An incorrectly sized embolization device may result in inadequate device placement, incomplete opening, or migration.All products are 100% inspected for damages and irregularities during manufacture.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received a report that at 3 month follow up, the flow diverter had migrated and was no longer covering the aneurysm neck.The patient is asymptomatic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX W/SHIELD TECHNOLOGY
Type of Device
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
COVIDIEN (IRVINE)
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
COVIDIEN (IRVINE)
9775 toledo way
irvine CA 92618
Manufacturer Contact
tricha miles
9775 toledo way
irvine, CA 92618
9492753836
MDR Report Key7041936
MDR Text Key93068812
Report Number2029214-2017-01272
Device Sequence Number1
Product Code OUT
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P100018.S011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/28/2019
Device Model NumberPED2-425-20
Device Lot NumberA351266
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/23/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age60 YR
-
-