• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CAPSTONE CONTROL SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CAPSTONE CONTROL SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Catalog Number 4011222
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/20/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.
 
Event or Problem Description
Pre-operative diagnosis for this procedure: lumbar canal stenosis type of procedure used: transforaminal lumbar interbody fusion levels implanted: l5-s1 it was reported that during surgery, inserter came off from a cage during rotation.When the cage was pulled out to confirm, it was found to be slightly deformed and the threads were broken.The cage was changed to one size smaller cage, and it was placed without problem.The product came in contact with the patient.No patient complications were reported as a result of the event.
 
Additional Manufacturer Narrative
Product analysis:the spacer was returned deformed from what appears to be a twisting motion.There is damaged noted at some points that interface with the inserter and the center section has collapsed from the twisting motion.This damage appears to be the result of torsional overload.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSTONE CONTROL SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key7042446
Report Number1030489-2017-02361
Device Sequence Number1916006
Product Code MAX
UDI-Device Identifier00643169109070
UDI-Public00643169109070
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K120368
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/04/2025
Device Catalogue Number4011222
Device Lot NumberH5356140
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2018
Initial Date Received by Manufacturer 10/20/2017
Supplement Date Received by Manufacturer03/22/2018
Initial Report FDA Received Date11/17/2017
Supplement Report FDA Received Date04/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age58 YR
-
-