• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC. SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM 3D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC. SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM 3D Back to Search Results
Model Number SDM-05000-3D3
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 10/20/2017
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that they were getting gridlines in the images causing the patient to be re-exposed.It was determined that the muliplier drawer needed to be replaced.Once this was done, the unit is working as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM 3D
Common Device Name
MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC.
36 & 37 apple ridge road
danbury CT 06810
MDR Report Key7042604
Report Number1220984-2017-00178
Device Sequence Number1399054
Product Code OTE
UDI-Device Identifier15420045505605
UDI-Public15420045505605
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
PO80003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Radiologic Technologist
Type of Report Initial
Report Date (Section B) 10/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Radiologic Technologist
Device Model NumberSDM-05000-3D3
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2017
Initial Report FDA Received Date11/17/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-