| Catalog Number |
60P050 |
| Medical Device Problem Code |
Torn Material (3024)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The tracheostomy tube is for single patient use, but can be used multiple times on the same patient.
|
| |
|
Event or Problem Description
|
|
It was reported that the tube was torn off when the patient was moving about while using a bivona® uncuffed pediatric tracheostomy tube.There were no patient adverse events.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was returned for evaluation without its original packaging.Visual inspection of the device found a split on the flange eyelet.A review of the testing and inspection documents was performed and deemed adequate and correct.A review of the manufacturing process was performed on a similar part and found no discrepancies.A review of 32 assemblies on the line was performed for molding issues and did not detect any damage.An attempt to reproduce the failure mode was performed by making a small cut on the inside and outside and pulled with a twill tape cut.Based on the evidence, a root cause was unable to be confirmed.It was determined that the most probable root cause is that damaged occurred after the device left the manufacturing facility.
|
| |
|