| Model Number |
2320 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Skin Tears (2516)
|
| Date of Event |
09/11/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned sensor was evaluated.The sensor failed external visual inspection, sliding neck tape at the sensor end caused exposed wire jackets and inner shield.The sensor passed continuity and functional testing.The wire jackets are exposed and not the wire itself, exposed jackets and inner shield do not carry any electrical current.
|
| |
|
Event or Problem Description
|
|
The customer reported a masimo lncs neo-3 spo2 pulse oximeter adhesive sensor from the nicu at iu north that was on a patient, the staff removed it from the patient when they noted minor harm due to the sensor having exposed wires.
|
| |
|