• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC PRECISION XCEED PRO; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC PRECISION XCEED PRO; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 70900-01
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/03/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product has been requested back for investigation.A follow- up report will be submitted once additional information is obtained.Note: the meter is designed to report readings between 20 mg/dl to 500 mg/dl.It should be noted that this meter does not give numeric value for readings less than 20 mg/dl.A "lo" display message indicates a reading less than 20 mg/dl.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
A caller reported receiving erratic readings on their adc blood glucose meter within 10 minutes.Results of 136 mg/dl and <20 mg/dl were plotted on the parkes error grid.The results fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury or mistreatment associated with this event.
 
Additional Manufacturer Narrative
The reported meter was returned and investigated with retained test strips.The complaint was not confirmed and no new issues were observed.All results were within range specification and no errors were observed during control solution testing.
 
Event or Problem Description
A caller reported receiving erratic readings on their adc blood glucose meter within 10 minutes.Results of 136 mg/dl and <20 mg/dl were plotted on the parkes error grid.The results fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury or mistreatment associated with this event.
 
Additional Manufacturer Narrative
Given no test strips were returned for investigation, retained test strip samples from the same lot reported by the customer (4c785h) were tested with control solution.All results were within the range specification and no errors were observed.The complaint is not confirmed.
 
Event or Problem Description
A caller reported receiving erratic readings on their adc blood glucose meter within 10 minutes.Results of 136 mg/dl and <20 mg/dl were plotted on the parkes error grid.The results fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION XCEED PRO
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key7043678
Report Number2954323-2017-08016
Device Sequence Number10792653
Product Code NBW
UDI-Device Identifier00093815709329
UDI-Public(01)00093815709329(17)20180831(10)4C785H
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K080960
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2018
Device Model Number70900-01
Device Catalogue Number70932
Device Lot Number4C785H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2017
Initial Date Received by Manufacturer 11/03/2017
Supplement Date Received by Manufacturer12/14/2017
01/10/2018
Initial Report FDA Received Date11/17/2017
Supplement Report FDA Received Date01/09/2018
01/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-