| Model Number |
70900-01 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/03/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been requested back for investigation.A follow- up report will be submitted once additional information is obtained.Note: the meter is designed to report readings between 20 mg/dl to 500 mg/dl.It should be noted that this meter does not give numeric value for readings less than 20 mg/dl.A "lo" display message indicates a reading less than 20 mg/dl.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
A caller reported receiving erratic readings on their adc blood glucose meter within 10 minutes.Results of 136 mg/dl and <20 mg/dl were plotted on the parkes error grid.The results fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported meter was returned and investigated with retained test strips.The complaint was not confirmed and no new issues were observed.All results were within range specification and no errors were observed during control solution testing.
|
| |
|
Event or Problem Description
|
|
A caller reported receiving erratic readings on their adc blood glucose meter within 10 minutes.Results of 136 mg/dl and <20 mg/dl were plotted on the parkes error grid.The results fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Given no test strips were returned for investigation, retained test strip samples from the same lot reported by the customer (4c785h) were tested with control solution.All results were within the range specification and no errors were observed.The complaint is not confirmed.
|
| |
|
Event or Problem Description
|
|
A caller reported receiving erratic readings on their adc blood glucose meter within 10 minutes.Results of 136 mg/dl and <20 mg/dl were plotted on the parkes error grid.The results fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury or mistreatment associated with this event.
|
| |
|
Search Alerts/Recalls
|