• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RFB LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RFB LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE Back to Search Results
Model Number 72200568F
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/03/2017
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the light source had a burnt resistor, resulting in overheating.No patient injury reported.Procedure completed with the same device.
 
Additional Manufacturer Narrative
There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the exterior of product and no physical damage was observed.A burnt resistor was found on the power supply.The light source failed power up.Cause of power failure is a defective power supply/ballast.Unit powered up and passed functional testing with a known good power supply installed.The complaint was confirmed and the investigation has concluded the cause of the failure to be a defective electronic component.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RFB LIGHT SOURCE 500XL XENON
Common Device Name
IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key7044280
Report Number1643264-2017-01740
Device Sequence Number1259999
Product Code FFS
UDI-Device Identifier00885556571392
UDI-Public(01)00885556571392
Combination Product (Y/N)N
PMA/510(K) Number
K994084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other,use
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72200568F
Device Catalogue Number72200568F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2017
Initial Date Received by Manufacturer 11/08/2017
Supplement Date Received by Manufacturer11/30/2017
Initial Report FDA Received Date11/19/2017
Supplement Report FDA Received Date12/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-